Tests for compatibility performed by customers
contribute greatly to our understanding of how
IRT/cs perform in some of the unusual environments
where non-contact temperature measurements are
required. CleveriR has a policy of providing sample
sensors with R&D discount for test purposes, in
return for a copy of test results. This series of tests
was performed by a customer who required the
performance, and had the facility to test the IRt/c.
The test was conducted in a 1.2 m by 1,.2 m (4 ft.
by 4 ft.) chamber in the following sequence:
First Test
1. Vacuum exposure (15 minutes at 40 Torr)
2. Microwave Exposure (5 minutes,
3 KW at 2450 MHz)
3. Chamber load (6.8 kg H2O)
Second Test
1. Vacuum exposure (10 minutes)
2. Microwave Exposure (30 seconds,
3 KW at 2450 MHz)
3. Microwave Exposure (10 seconds,
3 KW at 2450 MHz)
4. Chamber load (0.47kg H20)
The IRt/c showed no ill effects and operated
flawlessly when checked after the test sequence.